Model 6220/6221 Reference Manual Sweeps 4-5
Return to Section 4 topics
Logarithmic staircase sweeps
This sweep is similar to the linear staircase sweep. The steps, however, are done
on a logarithmic scale as shown in the example sweep in Figure 4-3. This exam-
ple is a 5-point log sweep from 1mA to 10mA. The sweep delay parameter deter-
mines the time period for each step. As with the linear staircase sweep, the sweep
delay period is the same for all steps.
Figure 4-3
Logarithmic staircase sweep (example 5-point sweep from 1mA to 10mA)
Delay
Delay
Delay
Delay
Start
Bias
Stop
(10)
Delay
Log Points = 5
1
10
5.6234
3.1623
1.7783
Current
(mA)
Log
Scale
Sweep delay is the
same for each step.
0