Model 6220/6221 Reference Manual Remote Operations 10-7
Return to Section 10 topics
Table 10-2
Remote interface configuration commands
Command Description
SYSTem:COMMunicate:SELect <name> Selects remote interface.
1
<name> = SERial, GPIB, or ETHernet.
GPIB interface command:
SYSTem:COMMunicate:GPIB:ADDRess <NRf> Sets GPIB primary address.
2
<NRf> = 0 to 30, default 12.
RS-232 interface commands:
SYSTem:COMMunicate:SERial:CONTrol:RTS <name> Selects handshaking.
3
<name> = IBFull, RFR, or OFF
SYSTem:COMMunicate:SERial:PACE <name> Sets flow control.
4
<name> = XON or OFF
SYSTem:COMMunicate:SERial:TERMinator <name> Sets output terminator.
5
<name> = CR, LF, CRLF, or LFCR.
SYSTem:COMMunicate:SERial:BAUD <n> Sets baud rate.
6
<n> = 300, 600, 1200, 2400, 4800,
9600, 19.2k, 38.4K, 57.6K, or 115.2K
SYSTem:COMMunicate:SERial:SEND <data> Sends data via RS-232.
7
SYSTem:COMMunicate:SERial:ENTer? Reads data from serial port.
8
SYSTem:LOCal Takes 622x out of remote.
9
SYSTem:REMote Puts 622x in remote.
9
SYSTem:RWLock <b> Enables or disables local lockout.
9
<b> = ON or OFF
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6220 and is the answer not in the manual?

Keithley 6220 Specifications

General IconGeneral
BrandKeithley
Model6220
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals