EasyManua.ls Logo

Keithley 6220

Keithley 6220
454 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Model 6220/6221 Reference Manual Wave Functions (6221 Only) 7-25
Return to Section 7 topics
Programming arbitrary waveforms
1. Restore defaults with this command:
*RST
2. Configure the waveform.
Examples – The following commands configure 100kHz arbitrary wave
with an amplitude of 25mA, 0mA offset, phase marker off, and 20 second
duration:
3. Arm and trigger the waveform, turn on output:
4. To stop generating the waveform and turn the source output off before the
duration elapses, send this command:
SOUR:WAVE:ARB:DATA -0.5, -0.25,
0, 0.3, 0.4
Define arbitrary data points in
location 0.
SOUR:WAVE:ARB:COPY 1 Copy points to location 1.
SOUR:WAVE:FUNC ARB1 Select arb wave, location 1.
SOUR:WAVE:FREQ 1e5 Set frequency to 100kHz.
SOUR:WAVE:AMPL 25e-3 Set amplitude to 25mA.
SOUR:WAVE:OFFS 0 Set offset to 0.
SOUR:WAVE:PMAR:STAT OFF Turn off phase marker.
SOUR:WAVE:DUR:TIME 20 20 second duration.
SOUR:WAVE:RANG BEST Select best fixed source range.
SOUR:WAVE:ARM ‘Arm waveform.
SOUR:WAVE:INIT Turn on output, trigger waveform.
SOUR:WAVE:ABOR Stop generating waveform.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

Table of Contents

Related product manuals