Model 6220/6221 Reference Manual Calibration 17-5
Return to Section 17 topics
Calibration summary and considerations
Calibration summary
Model 622x calibration procedures include:
• Front panel calibration, page 17-5
• Remote calibration, page 17-14
Calibration considerations
When performing the procedures:
• Be sure to restore factory defaults as outlined above.
• Make sure that the test equipment is properly warmed up and connected to
the correct Model 622x connector.
• Be sure that a shorting connector is connected to the INTERLOCK connec-
tor, or you will not be able to turn on the output for calibration. See Section
2 for details on the interlock.
• Make sure the test equipment and Model 622x calibration settings are
correct.
Calibration procedures
Front panel calibration
Using the calibration menu
Table 17-2 summarizes the main calibration menu. To enter the menu, press the
MENU key, select CAL, then press ENTER. Note that the calibration menus for the
Models 6220 and 6221 are almost identical except that the Model 6221 menu has
a WAVE calibration selection to calibrate the waveform generator while the Model
6220 does not.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6220 and is the answer not in the manual?

Keithley 6220 Specifications

General IconGeneral
BrandKeithley
Model6220
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals