Model 6220/6221 Reference Manual Triggering 8-3
Return to Section 8 topics
Remote trigger model
The trigger model for remote operation is shown in Figure 8-2.
Figure 8-2
Trigger model for remote operation
Arm-In
Event
Arm Event
Detector
Trigger Event
Detector
Device
Action
No
Ye s
Idle
Another
Arm
?
Another
Trigger
?
Ye s
No
Arm
Layer
Trigger
Layer
= Output Trigger
Trigger-In
Source
No
See Note
INITiate
?
Ye s
SOURce
ARM
:DIRection
ACCeptor
ARM:SOURce IMMediate
ARM:SOURce BUS
ARM:SOURce TIMer
ARM:SOURce MANual
ARM:SOURce TLINk
ARM:SOURce NSTest
ARM:SOURce PSTest
ARM:SOURce BSTest
TRIGger
:DIRection
SOURce
ACCeptor
TRIGger:OUTPut
SOURce | DELay | NONE
ARM:OUTPut
TEXit | NONE
SOURce:SWEep:COUNt <n>
TRIGger:SOURce IMMediate
TRIGger:SOURce TLINk
Note:
The following commands place the
Model 622x into idle: ABORt,
*RST, SYSTem:PRESet, *RCL <NRf>,
DCL, and SDC.
(Controls
Sweep
Steps)
(Controls
Entire
Sweep)
TRIGger:SIGNal
ARM:SIGNal
+
+
+
+
+
+
1
+
= Defaults
Sweep Parameters
Determine Number
of Steps
+
Sweep Step Occurs
During Device Action
ARM:OUTPut
TENTer | NONE
+
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6220 and is the answer not in the manual?

Keithley 6220 Specifications

General IconGeneral
BrandKeithley
Model6220
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals