Model 6220/6221 Reference Manual Delta, Pulse Delta, and Differential Conductance 5-63
Return to Section 5 topics
7. When finished with Differential Conductance, it can be disarmed by send-
ing the following command:
8. Recall stored Differential Conductance readings – Model 2182/2182A
Differential Conductance readings were sent to the buffer of the Model
622x. Send the following read command to read the buffer:
NOTE If a read command is sent when Differential Con-
ductance is not running, error -221 Settings Con-
flict will occur. See Table 5-2 for details on errors
associated with Differential Conductance opera-
tion.
NOTE Details on using Model 622x “Read commands”
are provided on page 5-15.
SOUR:SWE:ABOR ‘
‘
Stops Delta and places the Model 2182/2182A in
the local mode.
TRACe:DATA? ‘
‘
Read Differential Conductance readings stored in
622x buffer.