MMEMory3:STORe:DATA:AUTO:FNAMe <file>
MMEMory3:STORe:DATA:AUTO:FNAMe?
Selects the default file name for automatic storing of the Digitizer readings (see MMEMory3:STORe:DATA:AUTO
command) on the USB storage device after measurement complete. The file name optionally includes the folder
name and the .dat file extension.
This SCPI is supported on the DAQM909A Digitizer module only.
Parameter Typical Return
<file> = Any valid file name with IEEE-488.2 definite-length block,
including the mass storage unit specifier. Default:
“USB:\DIGAUTOSTORE.dat”
“USB:\MyDigAutoStore.dat”
Selects the “MyDigAutoStore.dat” on the USB storage device as the default file name:
MMEM3:STOR:DATA:AUTO:FNAM "USB:\MyDigAutoStore"
–
This file name will be saved in non-volatile memory on the instrument and will not be affected by a Factory
Reset (*RST command) or Instrument Preset (SYSTem:PRESet command).
–
The data only can store in binary format. If a .dat file extension is specified, or if no file extension is spe-
cified, the data is stored in binary format.
–
The destination folder must exist and cannot be marked as hidden or system.
–
If the destination file exists, it will be overwritten, unless marked as hidden or system.
–
This file name is set to its default value when the instrument is shipped from the factory and after
SYSTem:SECurity:IMMediate.