MMEMory3:STORe:DATA RDG_STORE, <file>
Stores all Digitizer measurements in reading memory to the specified data file. The file name optionally includes the
folder name and the .dat file extension.
This SCPI is supported on the DAQM909A Digitizer module only.
Parameter Typical Return
<file> = Any valid file name with IEEE-488.2 definite-length
block.
(none)
Stores all Digitizer measurements in reading memory to MyDigDAQ.dat in the root folder of the external USB mass storage file system:
MMEM3:STOR:DATA RDG_STORE,"USB:\MyDigDAQ"
The data only can store to a USB mass storage drive and in binary format. If a .dat file extension is specified, or if
no file extension is specified, the data is stored in binary format.
The destination folder must exist and cannot be marked as hidden or system.
If the destination file exists, it will be overwritten, unless marked as hidden or system.
MMEMory3:STORe:DATA:AUTO {OFF | ON}
MMEMory3:STORe:DATA:AUTO?
Enables (On) or disables (Off) the automatic storing of the Digitizer readings to a USB drive connected to the front
panel USB host port after measurement complete.
This SCPI is supported on the DAQM909A Digitizer module only.
Parameter Typical Return
{OFF | ON}
Default: OFF
0 (OFF) or 1 (ON)
Digitizer readings are to be stored to the USB drive after measurement complete: MMEM3:STOR:DATA:AUTO ON
This setting will be saved in non-volatile memory on the instrument and will not be affected by a Factory Reset
(*RST command) or Instrument Preset (SYSTem:PRESet command).
When USB automatic storing is enabled, readings are saved to file on the USB drive inserted in the front panel
USB host port after measurement complete.
If you remove the USB drive during storing, storing will stop. If you re-attach the USB drive, it will not cause stor-
ing to resume. However, readings are still able to store manually using MMEMory3:STORe:DATA command.
This setting is set to its default value when the instrument is shipped from the factory and after
SYSTem:SECurity:IMMediate.
196
Keysight DAQ970A/DAQ973A Programming Guide
2 SCPI Programming

Table of Contents

Other manuals for Keysight Technologies DAQ970A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies DAQ970A and is the answer not in the manual?

Keysight Technologies DAQ970A Specifications

General IconGeneral
BrandKeysight Technologies
ModelDAQ970A
CategoryMultiplexer
LanguageEnglish