[SENSe:]FUNCtion[:ON] "<function>" [, (@<ch_list>)]
[SENSe:]FUNCtion[:ON]? [(@<ch_list>)]
Only supported for all the measurement functions on DAQM900A, DAQM901A, DAQM902A, and DAQM908A
multiplexer modules.
This command is not supported in DMM Digitize data acquisition mode.
Error -221,"Settings conflict; DMM digitizing mode has been disabled" will be generated when you send the
command to change measurement function in DMM Digitize mode.
Selects the measurement function on the selected channels (all function-related measurement attributes are
retained).
Parameter Typical return
CAPacitance
CURRent:AC
CURRent[:DC]
DIODe
FREQuency
PERiod
FRESistance
RESistance
STRain:DIRect
STRain:FDIRect
STRain:QUARter
STRain:HALF:BENDing
STRain:HALF:POISson
STRain:FULL:BENDing
STRain:FULL:BENDing:POISson
STRain:FULL:POISson
TEMPerature[:TC]
TEMPerature:FRTD
TEMPerature:RTD
TEMPerature:FTHermistor
TEMPerature:THERmistor
VOLTage:AC
VOLTage[:DC]
Default: VOLTage[:DC]
The short form of the selected function is returned in quotation marks, with no optional keywords:
"CAP", "CURR:AC", "DIOD", "TEMP", and so on.
Selects the AC voltage function: FUNC "VOLT:AC"
– The function name must be enclosed in quotes in the command string.
– If you change the measurement function, all measurement attributes of the previous function (range, resolution,
etc) are returned to the default values. If you return to the original function, those attributes are restored.
Changing the measurement function disables scaling and limit testing (CALC:SCAL:STAT and CALC:LIM:STAT set
to OFF).
– To change measurement function in DMM Digitize mode, use the ACQuire:<function> family command.
– This parameter is set to its default value after a Factory Reset (*RST command).
224
Keysight DAQ970A/DAQ973A Programming Guide
2 SCPI Programming