SYSTem:SECurity:IMMediate sets this parameter to its default value.
A value of "0.0.0.0" or "255.255.255.255" indicates that subnetting is not being used.
The setting is non-volatile, and does not change when power has been off, after a Factory Reset (*RST com-
mand), or after an Instrument Preset (SYSTem:PRESet command).
SYSTem:COMMunicate:LAN:TELNet:PROMpt "<string>"
SYSTem:COMMunicate:LAN:TELNet:PROMpt?
Sets the command prompt displayed when communicating the instrument with Telnet.
Parameter Typical return
<string> A string of up to 15 characters "Command"
Defines the command prompt: SYST:COMM:LAN:TELN:PROM "Command"
Query returns the command prompt as ASCII strings enclosed in double quotes.
Instrument uses LAN port 5024 for SCPI Telnet sessions, and port 5025 for SCPI Socket sessions.
Telnet port is an alternate way to send SCPI commands to the instrument.
Telnet session can typically be started as follows from a host computer shell:
telnet <IP_address> <port>
For example:
telnet 169.254.4.10 5024
To exit a Telnet session, press <Ctrl-D>.
SYSTem:SECurity:IMMediate sets this parameter to its default value.
The setting is non-volatile, and does not change when power has been off, after a Factory Reset (*RST com-
mand), or after an Instrument Preset (SYSTem:PRESet command).
332
Keysight DAQ970A/DAQ973A Programming Guide
2 SCPI Programming

Table of Contents

Other manuals for Keysight Technologies DAQ970A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies DAQ970A and is the answer not in the manual?

Keysight Technologies DAQ970A Specifications

General IconGeneral
BrandKeysight Technologies
ModelDAQ970A
CategoryMultiplexer
LanguageEnglish