– SYSTem:SECurity:IMMediate sets this parameter to its default value.
– A value of "0.0.0.0" or "255.255.255.255" indicates that subnetting is not being used.
– The setting is non-volatile, and does not change when power has been off, after a Factory Reset (*RST com-
mand), or after an Instrument Preset (SYSTem:PRESet command).
SYSTem:COMMunicate:LAN:TELNet:PROMpt "<string>"
SYSTem:COMMunicate:LAN:TELNet:PROMpt?
Sets the command prompt displayed when communicating the instrument with Telnet.
Parameter Typical return
<string> A string of up to 15 characters "Command"
Defines the command prompt: SYST:COMM:LAN:TELN:PROM "Command"
– Query returns the command prompt as ASCII strings enclosed in double quotes.
– Instrument uses LAN port 5024 for SCPI Telnet sessions, and port 5025 for SCPI Socket sessions.
– Telnet port is an alternate way to send SCPI commands to the instrument.
– Telnet session can typically be started as follows from a host computer shell:
telnet <IP_address> <port>
For example:
telnet 169.254.4.10 5024
To exit a Telnet session, press <Ctrl-D>.
– SYSTem:SECurity:IMMediate sets this parameter to its default value.
– The setting is non-volatile, and does not change when power has been off, after a Factory Reset (*RST com-
mand), or after an Instrument Preset (SYSTem:PRESet command).
332
Keysight DAQ970A/DAQ973A Programming Guide
2 SCPI Programming