CALibration:TIME?
Returns the amount of time since the last calibration in the format hh,mm,ss.sss.
Parameter Typical return
(none) 20,15,30.000
Returns the calibration time: CAL:TIME?
You can perform this query regardless of whether the instrument is secured.
This setting is non-volatile. It will not be changed by power cycling, after a Factory Reset (*RST command) or
Instrument Preset (SYSTem:PRESet command).
CALibration:VALue <value>
CALibration:VALue?
Specifies the value of the applied calibration signal.
Parameter Typical return
<value> = Numeric value.
Default: 0.0
+1.00101000E+01
Sets calibration value to +10.001010 volts: CAL:VAL10.001010
This setting is non-volatile. It will not be changed by power cycling, after a Factory Reset (*RST command) or
Instrument Preset (SYSTem:PRESet command).
Keysight DAQ970A/DAQ973A Programming Guide 87