E8257D/67D, E8663D PSG Signal Generators Service Guide 417
Post–Repair Procedures
CPU Configuration
— If the license key(s) for your specific instrument is not available from
the GSDC website, send a message to the PSG,Sources email node
requesting the license keys along with the following information:
— Instrument model number
— Serial number
— Host ID
— A list of the options
The request will be researched, and if the database indicates that
the list of options were installed in the instrument, license keys will
be sent. If the database does not show that an option(s) was
installed, no license key will be sent for that option. License keys
will only be provided for verified options.
12.Install the license keys:
a. Press Utility > Instrument Adjustments > Instrument Options >
Software Options.
b. Using the RPG knob, highlight the desired option.
c. Press Modify License Key.
d. Using the numeric key pad and the softkeys, enter the license key.
When the signal generator accepts the license key, a check mark
appears next to the license
e. After entering all the required license keys, press Proceed With
Reconfiguration.
If a license key does not turn an option on, proceed to Step 11 and
perform the step as if the license keys had not been retrievable from
the CPU.