3 Getting Started
58 Keysight N6700 User’s Guide
4 For additional help about any of the pages, click on the Help with
this Page button.
If desired, you can control access to the Web server using password
protection. As shipped from the factory, no password is set. To set a
password, click on the View & Modify Configuration button. Refer to the
on-line help for additional information about setting a password.
In an MS-DOS Command Prompt box type: telnet hostname 5024 where
hostname is the N6700 hostname or IP address, and 5024 is the
instrument’s telnet port.
You should get a Telnet session box with a title indicating that you are
connected to the power system. Type the SCPI commands at the
prompt.
Power system mainframes allow any combination of up to
simultaneous
data socket, control socket, and telnet connections to be made.
Keysight instruments have standardized on using port 5025 for SCPI
socket services. A
on this port can be used to send and
receive ASCII/SCPI commands, queries, and query responses. All
commands must be terminated with a newline for the message to be
parsed. All query responses will also be terminated with a newline.
The socket programming interface also allows a
connection. The control socket can be used by a client to send device
clear and to receive service requests. Unlike the data socket, which uses
a fixed port number, the port number for a control socket varies and
must be obtained by sending the following SCPI query to the data
socket: SYSTem:COMMunicate:TCPip:CONTrol?
After the port number is obtained, a control socket connection can be
opened. As with the data socket, all commands to the control socket
must be terminated with a newline, and all query responses returned on
the control socket will be terminated with a newline.
To send a device clear, send the string “DCL” to the control socket.
When the power system has finished performing the device clear it
echoes the string “DCL” back to the control socket.
Service requests are enabled for control sockets using the Service
Request Enable register. Once service requests have been enabled, the
client program listens on the control connection. When SRQ goes true
the instrument will send the string “SRQ +nn” to the client. The “nn” is
the status byte value, which the client can use to determine the source
of the service request.