SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
SA Mode Settings
The individual alignments may be:
— Coupled (default) - when they are in the same state, or
— UnCoupled - when at least one of them is in a different state than the other 
ones 
— Executed on demand, all of them, using the Align All Now
 softkey.
Each of the individual alignments:
— could be in one of three states:
— Off - the alignment is not applied and is not updated automatically
— Auto (default) - the alignment will be updated over time and 
temperature as needed, no user interaction required
— Hold - the unit holds the alignment values from the last 
alignment-execution and does not execute another alignment 
automatically
— can be updated on demand using the corresponding Align Now
 soft-key. 
Refer to “Individual Alignments” on page 147.
—The All Alignment
 soft-key has the following functionality:
— when the individual alignments are Coupled, the state of all of them 
can be changed simultaneously using the All Alignment
 soft-key 
sub-menus
— when the individual alignments are UnCoupled, they will be coupled 
when using the All Alignment
 soft-key sub-menus and their state 
will be simultaneously changed accordingly
All Alignment and Align All Now
IMPORTANT! It is recommended that you use Align All Now, because it 
includes the individual alignment InstAlign Amptd Align Now
, as well as, 
Burst Align Now
, and ChanEQ Align Now. But, if you would prefer to have 
manual control over the individual alignments, then refer to the SA 
Individual Alignment Control setting in “Preferences” on page 230, in 
Chapter 9 System Settings.
The notification AMPTD? displayed on screen means the InstAlign amplitude alignment or 
channel equalization is stale (if its state is Hold) and requires an update or it is in the OFF state. 
This situation does not occur when the respective alignment is in the Auto state. To clear this 
notification, press the 
Align All Now softkey.