Keysight N9938-90003 User’s Guide 53
DTF (Distance to Fault) Measurements
How to make DTF Measurements
9. Either press Edit/Save/Recall Cables then Recall Cable, or directly enter
the Velocity Factor/Cable Loss of the DUT using the Velocity Facto
r and
Cable Loss
keys.
10. Connect the start end of the DUT to the FieldFox.
11. Press Meas Setup 4
then Settings then Next Page. If the Alias-free Range
setting is False, then you may see Alias faults on the screen. Learn more
on page 61.