Keysight N774xC Series Multiport Power Meter Programming Guide 29
Instrument Setup and Status 3
1Not Used 2
0 (LSB) Operation Complete 1
Response: none
Example: *ESE 255
Command: *ESE?
Syntax: *ESE?
Description: The standard Event Status Enable query *ESE? returns the contents of the Standard Event Status Enable Mask (see *ESE
for information on this register).
Parameters: none
Response: The bit value for the register (a 8-bit integer value).
Example: *ESE? −> +255
Command: *ESR?
Syntax: *ESR?
Description: The standard Event Status Register query *ESR? returns the contents of the Standard Event Status Register. The register
is cleared after being read.
parameters none
response The bit value for the register (a 8-bit integer value):
Bit Mnemonic Decimal Value
7 (MSB) Power On 128
6Not used 64
5 Command Error 32
4 Execution Error 16
3 Device Dependent Error 8
2Query Error 4

Table of Contents