82 Keysight N774xC Series Multiport Power Meter Programming Guide
4 Measurement Operations & Settings
Response: None
Example: :sens1:pow:gain:auto 1
Affects All power meters
Command: :SENSe[n]:POWer:GAIN:AUTO?
Syntax: :SENSe[n]:POWer:GAIN:AUTO?
Description: Get the Auto Gain.
Parameters: None
Response: 0 = Auto Gain Off.
This is the position for best transient response.
1 = Auto Gain On (Default)
This is the Position for best dynamic.
Example: :sens1:pow:gain:auto? 1
Affects All power meters
Command: :SENSe[n]:POWer:ATIMe
Syntax: :SENSe[n]:POWer:ATIMe<wsp><averaging time>[NS|US|MS|S]
Description: Sets the averaging time.
Parameters: The averaging time as a float value in seconds.
If you specify no units in your command, seconds are used as the default.
NONE: For N774xC power meters the internal granularity of the averaging time is 1us which allows flexibility.
Response: None
Example: :sens1:pow:atim 1s
Affects All power meters
Command: :SENSe[n]:POWer:ATIMe?
Syntax: :SENSe[n]:POWer:ATIMe?
Description: Returns the averaging time.
Parameters: None