Keysight N774xC Series Multiport Power Meter Programming Guide 83
Measurement Operations & Settings 4
Response: The averaging time as a float value in seconds.
Example: :sens1:pow:atim? → +1.00000000E+000
Affects All power meters
Command: :SENSe[n]:POWer:RANGe:AUTO
Syntax: :SENSe[n]:POWer:RANGe:AUTO <wsp><boolean>
Description: Enables or disables automatic power ranging for the channel.
If automatic power ranging is enabled, ranging is automatically determined by the instrument. Otherwise, it must be set by
the sens[n]:pow:rang command.
Automatic ranging while other commands are sent to power meters has lead to timing conflicts in some configurations.
Automation programs can often better control the range directly.
Parameters: A boolean value: 0 or OFF: automatic ranging disabled
1 or ON: automatic ranging enabled
Response: None
Example: :sens1:pow:rang:auto 1
Affects All power meters
Command: :SENSe[n]:POWer:RANGe:AUTO?
Syntax: :SENSe[n]:POWer:RANGe:AUTO?
Description: Returns whether automatic power ranging is being used by the channel.
Parameters: None
Response: A boolean value: 0 or OFF: automatic ranging disabled
1 or ON: automatic ranging enabled
Example: :sens1:pow:rang:auto? → 1
Affects All power meters