EasyManua.ls Logo

Keysight N774 C Series - Page 98

Keysight N774 C Series
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
98 Keysight N774xC Series Multiport Power Meter Programming Guide
5 Error Codes
-150 “String data error”
[This error, as well as errors -151 through -159, are generated when parsing a string data element.
This error message is used when the device cannot detect a more specific error.]
-151 “Invalid string data
[A string data element was expected, but was invalid for some reason (see IEEE 488.2, 7.7.5.2); for
example, an END message was received before the terminal quote character.]
-158 “String data not allowed”
[A string data element was encountered but was not allowed by the device at this point in parsing.]
-160 "Block data error"
-161 “Invalid block data”
[A block data element was expected, but was invalid for some reason (see IEEE 488.2, 7.7.6.2); for
example, an END message was received before the length was satisfied.]
-168 “Block data not allowed”
[A legal block data element was encountered but was not allowed by the device at this point in
parsing.]
-170 “Expression error”
[This error, as well as errors -171 through -179, are generated when parsing an expression data
element. This particular error message is used when the device cannot detect a more specific
error.]
-171 “Invalid expression”
[The expression data element was invalid (see IEEE 488.2, 7.7.7.2); for example, unmatched
parentheses or an illegal character.]
-178 “Expression data not allowed”
[A legal expression data was encountered but was not allowed by the device at this point in
parsing.]
-180 "Macro error"
-181 “Invalid outside macro definition”
[Indicates that a macro parameter placeholder ($<number) was encountered outside of a macro
definition.]
-183 “Invalid inside macro definition
[Indicates that the program message unit sequence, sent with a *DDT or *DMC command, is
syntactically invalid (see IEEE 488.2, 10.7.6.3).]
-184 "Macro parameter error"
Error
Number String

Table of Contents

Other manuals for Keysight N774 C Series