This instrument includes 11 calibration channels from Ch00 to Ch10.
Ch00 is for measurement based upon KONICA MINOLTAs calibration standard. Its calibration correction
coefficient has been set and is unchangeable.
The following contents can be set to Ch01 to Ch10 respectively using CS-S30 software for spectroradi-
ometers included in the instrument’s standard accessories. For details on how to set up using the soft-
ware, see the CS-S30 instruction manual.
Correction coefficient of user calibration
Correction coefficient ID
They are commonly used among each color space of Lvxy, Lvu’v’, LvTcpΔuv, XYZ, dominant wavelength/
excitation purity and spectral graph in one channel.
Calibration channels can be changed using the following procedure.
Operating Procedure
1
6, 7
1. Press
MENU
key when the MEAS screen
is displayed.
The MENU screen appears on the LCD.
When the backlight of the LCD has been turned off via
BACKLIGHT
key on the MEAS screen, the
backlight is turned on.
MEAS SNGL
UC00
SPD
Nrm
CMF
2
˚
SYN
[
Hz
]
59.94
ACC
Non
cd
m
2
Lv
63.60
0 . 4015
0 . 4061
x
y
1
˚
MEAS
MEMORY
OPTION
SETUP
DARK MEASUREMENT
INFORMATION
MENU
2, 3, 4, 5