Routine oxygen sensor replacement required.
Order replacement kit from www.korr.com. Follow
replacement instructions as indicated in the
Maintenance section of this manual. Restart test.
No oxygen sensor
connected.
Oxygen sensor replacement required. Order
replacement kit from www.korr.com. Follow
replacement instructions as indicated in the
Maintenance section of this manual. Restart test.
Oxygen Sensor
older than 18
months.
Acquire new oxygen sensor soon from
www.korr.com.
Oxygen sensor
reading too high.
Allow unit to sit motionless for a time then try
running test again. If error persists, contact KORR at
www.korr.com/support/service to arrange for
service at the nearest KORR authorized service
center.
Oxygen sensor
reading unstable.
Allow unit to sit motionless for a time then try
running test again. If error persists, contact KORR at
www.korr.com/support/service to arrange for
service at the nearest KORR authorized service
center.
RMR testing: Try using a different
MetaBreatherMAX in case of a faulty valve.
VO
2
Max Testing: Was the Y-valve taken apart
recently? Check the diaphragm. It may be in
backwards, or not nested in correctly. Reference
this video for instructions regarding reassembling
the Y-valve properly:
https://youtu.be/Ve5PkfMMpTk
VO
2
readings too
low during test.
Test will end, or
abort unless
corrected quickly.
If error pops up prior to test, have the client take
five-six big deep breaths. This will force air to
accumulate in the mixing chamber and trigger the
90 second countdown to start.
If error pops up mid test, volume of air measured is
too low. May be due to a leak in the accessories.
Check for the following:
● Good seal around MetaBreatherMAX
mouthpiece
● Mask fits snuggly
● Y-Valve assembled correctly