1 PRODUCT FEATURES
10
OPTISONIC 6300
www.krohne.com 09/2020 - 4006413102 - TD OPTISONIC 6300 V2 R02 en
X mode
For an easy installation, optimal accuracy, maximum reliability and reducing risk: X-mode
X-modeX-mode
X-mode
By positioning two rails opposite each other on the pipe a dual, direct path solution is
established. This provides the following advantages:
A direct path without reflection lowers measurement uncertainty and risk on path loss.
Dual path provides redundancy. A failing path is automatically compensated by the dynamic
path replacement.
Diagnostic features
By featuring several diagnostic options the quality of the measurement can be monitored over
time. Signal quality parameters like; signal to noise, signal strength and stability are available
for monitoring changes in process, fluid properties, product change, empty pipe, cavitation and
so on. This allows for condition-based maintenance, keeping the flowmeter in optimal condition
and avoiding unplanned downtime. Diagnostics are in full compliance with NAMUR NE 107
available through the communication interfaces or shown in the display.
Figure 1-6: X beam configuration of medium version
Failure
Output signal invalid
Check function
Output signal (temporarily) invalid
Out of specification
Unreliability of output signal
Maintenance required
Output signal still valid
UFC 300 diagnostic capabilities: NE107
UFC 300 diagnostic capabilities: NE107UFC 300 diagnostic capabilities: NE107
UFC 300 diagnostic capabilities: NE107
NE107 icons for status messages and error
handling
- visible on UFC 300 display
- via all communication protocols
- Status messages are grouped by problem
source
- User can change group or priority