INSTALLATION 3
41
OPTISONIC 6300
www.krohne.com09/2020 - 4006413102 - TD OPTISONIC 6300 V2 R02 en
3.16.3 Turning the display of the field housing version
The display of the field housing version can be turned in 90° increments
1 Unscrew the cover from the display and operation control unit.
2 Using a suitable tool, pull out the two metal puller devices to the left and right of the display.
3 Pull out the display between the two metal puller devices and rotate it to the required position.
4 Slide the display and then the metal puller devices back into the housing.
5 Re-fit the cover and tighten it by hand.
Figure 3-18: Turning the display of the field housing version
The ribbon cable of the display must not be folded or twisted repeatedly.
Each time a housing cover is opened, the thread should be cleaned and greased. Use only resin-
free and acid-free grease.
Ensure that the housing gasket is properly fitted, clean and undamaged.