37 User Manual - LCR Pro1/Pro1 Plus - ver 1.06
Characteristics and Specifications
Product Characteristics
Table 4-1 Product Characteristics
Dimensions (L x W x H)
151 x 19 x 15mm
Weight
30 grams
Display
0.96-inch, 128x64 OLED display
Battery
150 mAH internal lithium-ion polymer battery
Battery Life
1 day in typical measurement
(1)
Charging Source
USB port
USB power adapter (output voltage DC 5V ± 5%)
Charging Time
2.5 hours (typical)
Measurement Speed
0.5sec, 1sec (typical) or 2 seconds
Operating Environment
Operating temperature from -10°C to 50°C, 0% to 80% RH
Full accuracy up to 80% RH for temperature 23°C ± 3°C
Storage Compliance
-20°C to 60°C, 0% to 80% RH
Safety and EMC
Compliance
IEC61000-4-2 - ESD (4 kV Contact, 8 kV Air)
EN 61000-4-3 - Radiated Immunity
IEC61000-4-8 - Magnetic Field Immunity
FCC15/EN 55011/ICES-003 - Class B, Radiated Emissions
FCC15 Class B Conducted Emissions
Calibration Cycle
1 Year (Recommended)
NOTE
Battery life varies by use, configuration, and many other factors. Actual results may vary.

Table of Contents

Other manuals for LCR Pro1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LCR Pro1 and is the answer not in the manual?

LCR Pro1 Specifications

General IconGeneral
BrandLCR
ModelPro1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals