EasyManua.ls Logo

Lightmed TruScan Pro - Page 40

Lightmed TruScan Pro
109 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LIGHTMED TruScan Pro Operator’s Manual Rev. No. 01 Page 40 of 109
Activated
When this button is pressed, the total laser
shot counter is reset to zero.
4
Shot counter
Displays the total laser shots fired during the treatment session.
5
Total Energy (J)
To indicate the total delivered energy to the patient, in J.
6
SP-Mode
Inactive
The sub-threshold pulse (SP) mode is
OFF.
Activated
The sub-threshold pulse (SP) mode is ON. The SP-
mode duty cycle is selectable from 2.5% to 50%.
7
Aiming beam
intensity
The aiming beam intensity can be adjusted with the two up and
down arrows.
8
Aiming beam
intensity display
Displays the brightness of the aiming beam. At the maximum
brightness, the output power is around 1.0 mW or less to the
patient’s cornea.
9
Software version
Displays the current software version.
10
Spot size on retina
Displays the spot size on patient’s retina in µm.
11
Spot size display
Displays the spot size in µm. The spot size quick menu can be
accessed by clicking on the right arrow to expand the following spot
size options:
50 µm
100 µm
200 µm
300 µm
400 µm
500 µm
12
Duration controls
The duration of laser exposure to the patient each time the
footswitch is pressed in the “READY” mode can be adjusted by the

Table of Contents

Related product manuals