-54-
■
External trigger output 2 (OT2 pulse output)
It outputs an about 1m second
L pulse to the external trigger
output 2 (OTS2).
From the trigger output 1 (OT1) it
outputs about 1m second L pulse
when the trigger factor
meets regardless of the trigger action setting.
→ 7.2 External Trigger Input/Output
7.4 Timer / counter function
The timer and counters used with the trigger function - the timer
(Timer0 / Timer1), counter (Counter0 / Counter1), and counter for
transmission/reception data (Counter TXD / RXD) are available.
Timer/Counter Action
■
Timer (Timer0 / Timer1) action
1) It is cleared to 0 when starts measurement and remains in stop status.
2) It actions depending on the trigger action by the trigger function –
start (0: Start), stop (1: Stop), or restart (2: Restart). When it actions
as Start or Restart, the current timer value counts up with the time
resolution set in advance.
3) The set value for comparison of timer 0,1 and the matching
information of the current value of timer 0,1 are used for the trigger
factor of timer matching.
4) When the current value overows, it continues the timer from 0.
5) If you stop the measurement, the timer stops at the moment.
■
Counter (Counter0 / Counter1) action
1) It is cleared to 0 when starts measurement.
2) By the trigger action of trigger function it increments (0: Increment)
or clears (1: Clear) the value.
3) The set value for comparison of counter 0,1 and the matching
information of the current value of counter 0,1 are used for the
trigger factor of counter matching.
4) When the current value overows, it continues counting from 0.