At the top there is placed a panel with colour indicator lamps. The lit letter A, B
or C means of the attending doctor´s option. Furthermore here is a green
indicator lamp indicating the connection of the chair to the network, an orange
indicator lamp indicating the connection of wireless control to the chair. The
function, which does not lit, is switched off. There is a yellow battery status
indicator on the left (see the Backup Battery Section) and a red indication of two
collision situations videocolposcope vs. leg supports and setting the chair on
the uneven floor and thus not closed security SLS switch.
.
Two minutes after the chair is unplugged from the power the chair goes into
sleep mode, the led lights go out and the remote control stops working. The
chair can be activated again by pressing any button on the hand or foot
controller.
Do not stand in the space under the motor driven parts of the loading area
(back, seat and footrest)
In case of the controller failure (permanent pushing the button or cable
damage), the drive can be stopped by pressing the STOP button on the
hand control. Right after that the supply connection must be disconnected
and a representative of the L I N E T spol. s r.o. Company must be
contacted.
In case of blocking of the motor motion we advise to simultaneously push
both upper buttons of the hand control (adjustment of the chair lifting)
acoustic signalling will sound for 5 seconds. During the acoustic signalling
we keep the buttons pushed. All saved positions will be deleted.
If the above mentioned does not help, we recommend to incline the back
part into maximum position (vertical position), to lower down the footrest
maximally and to bring the chair to the lowest position. Then pull the energy
supply cable out for ten seconds.
The gynaecological chair is not intended for permanent operation
motors can be burdened maximally to the values of 2 minutes
operation and then 18 minutes idle state. In case of overburden the
operation can be automatically stopped. Service action must be carried
out please, contact a representative of the L I N E T spol. s r.o.