20.1 Cleaning (Multicare X)
Prepare for cleaning as follows:
Put the mattress support platform in the highest position.
Adjust the back and thigh rests so that the reverse sides are accessible.
Disable the function buttons on the control elements using the Attendant Control Panel.
Disable the foot controls using the Attendant Control Panel.
Disconnect the bed from the mains.
Move the bed to the location where it will be cleaned.
Lock the brakes on the bed.
20.1.1 Daily Cleaning
Clean the following bed parts:
All control elements for adjusting the bed
All handles
   □ CPR release handle
Head board and Foot board
Siderails (in highest position)
■  Freely accessible mattress surface
Mobi-Lift
®
Accessory rails
20.1.2 Cleaning before Changing Patients
Clean the following bed parts:
All control elements for adjusting the bed
All handles
   □ CPR release handle
Head board and Foot board
Siderails (in highest position)
Freely accessible mattress surface
Mobi-Lift
®
Accessory rails
All plastic mattress support platform covers
Plastic undercarriage covers
Telescopic columns
Mattress on all sides
Freely accessible metal parts of mattress support platform
Cable ducts
Lifting pole sleeve tting
Infusion stand sleeve tting
Bumpers
Castors
Brakes
For safe and gentle cleaning:
Do not use any strong acids or bases (optimum pH range 6 - 8).
Exclusively use detergents that are suitable for cleaning medical equipment.
Do not use abrasive powders, steel wool, or other materials and cleaning agents that might damage the mattress replace-
ment system.
Never use any corrosive or caustic detergents.
Never use detergents that deposit calcium carbonate.
Never use detergents with solvents that might a󰀨ect the structure and consistency of the plastics (benzene, toluene, ace-
tone, etc.).
Clean electrical components carefully and allow them to dry completely.
Do not immerse SCU in water or steam-clean it.
Observe local directives regarding infection control.
Make sure any cleaning agent used is approved by:
   ■ the facility in which the mattress replacement system is to be used.
   ■ by the environmental protection agency of the country in which the mattress replacement system is to be used.