6 Use and Storage Conditions
DANGER!
Danger to life due to electric shock!
To ensure the bed’s class I protection against electric shocks:
Ground the mains.
Use exclusively Hospital Grade or Hospital Only receptacles for grounding.
Multicare X and OptiCare X are designed for use in rooms for medical purposes. Electrical installations must therefore meet local
norms laying down the necessary conditions for electrical installations.
Disconnect the bed from the mains in exceptional cases (i.e. lightnings, earthquake).
Multicare X and OptiCare X are not suitable for indoor environments containing ammable gases (except oxygen cylin-
ders). Respect values of the parameters connected with environment conditions in the chapter Technical Specication 
during use and storage of the product.