D9U001MCX-0101_03
36
9 Power Cord
Attachment plug is means of connecting and disconnecting bed from the mains.
Mains power cord must be attached with a hook at the head end of the bed during transport.
CAUTION!
Disconnecting bed from the mains does not stop motions of the bed!
► Stop the bed before disconnection bed from the mains.
Where the integrity of the external protective conductor in the installation or its arran-
gement is in doubt
► operate the bed from internal accumulator only.
Fig. Hook for hanging power cord
8.8 Firmware
The bed includes rmware that can be updated only by an authorised service technician.
This rmware is protected against unauthorised access by mechanical housing (tool is needed to access), by seal (com-
ponents with processor are sealed), by exclusive compatibility with an authorised software tool and by check of compati-
bility of the new rmware with the bed.
Fig. Bed Transport Mode (Attendant Control Panel)

Table of Contents