12.1.3 Fault Codes (OptiCare X integrated mattress)
Fault Code
Type of the Fault Immediate Action of Operator Corrective Action of Operator
P2
Major Inate Fault
(air leak)
Check both manual CPR valves for closing.
OR
Check air connections to SCU.
If problem is still present, transfer patient to
another surface.
If problem is still present,
contact service department
approved by manufacturer!
P3 Deate Fault
Wait three minutes until fault self-clear.
OR
Check that air pipes between mattress and SCU
are not kinked.
If problem is still present, transfer patient to
another surface.
If problem is still present,
contact service department
approved by manufacturer!
P4 Pressure Sensor Fault Transfer patient to another surface.
Contact service department
approved by manufacturer!
P5 Overpressure Fault Transfer patient to another surface.
Contact service department
approved by manufacturer!
P6 BOD Valve Fault
If the mattress is still supporting the patient, they
should not move the patient to another surface if
fault self-clears until 60 minutes.
OR
Transfer patient to another surface.
Contact service department
approved by manufacturer!
P7 Blower Fault No MCM function, assess the e󰀨ect on patient.
Contact service department
approved by manufacturer!
P8 Clogged Blower Filter No MCM function, assess the e󰀨ect on patient.
Contact service department
approved by manufacturer!
P9 Communication Fault Transfer patient to another surface.
Contact service department
approved by manufacturer!
P12 Optimization Out of Range
Patient may be too heavy for the mattress or
in a position that could prevent the BOD from
closing. Check patient weight, patient‘s position
and other medical device that is being used (e.g.
traction, positioning aids etc.).
OR
Transfer patient to another surface.
Contact service department
approved by manufacturer!
P14 Overcurrent of Solenoid Transfer patient to another surface.
Contact service department
approved by manufacturer!
P92
System Control Unit
Overheating
Transfer patient to another surface.
Contact service department
approved by manufacturer!
P94 Pressure Calibration Fault Transfer patient to another surface.
Contact service department
approved by manufacturer!
P98 Solenoid Fail Transfer patient to another surface.
Contact service department
approved by manufacturer!
P100
Find the corresponding graphic pop-up in the chapter „12.1.2 Statuses (Multiboard X - LCD touchscreen)“ on
the page 51 and follow the instructions related to it.
P101
P102
P105
P106
P107
P108