EasyManua.ls Logo

LINET Multicare X

LINET Multicare X
169 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D9U001MCX-0101_03
90
13.12 Lateral Tilt
It is not possible to position Lateral Tilt with a siderail folded down without use of an additional function.
To position Lateral Tilt use:
► Multiboard X
► Lateral Tilt Foot Control
Attendant Control Panel
Multiboard X Display shows Lateral Tilt Angle. Maximum Lateral Tilt Angle is 30 degrees.
Fig. Lateral Tilt Angle on
Multiboard X Display
The Maximum Lateral Tilt Angle depends on position of the mattress support platform, on position of siderails (Transfer
Mode), on how much the bed is loaded and on patient position on the bed.
The following table shows Maximum Possible Lateral Tilt Angle values in relation to the patient’s weight, while lying in the
middle of the bed. Other position of the patient in the bed can decrease the Maximum Lateral Tilt Angle.
Maximum Possible Lateral Tilt Angle Working Load
250 kg
10° 225 kg
15° 199,5 kg
20° 170 kg
25° 153 kg
30° 150 kg

Table of Contents

Related product manuals