D9U001MCX-0101_03
24
Parameter Value
Input Voltage, Frequency 230 V AC, 50/60 Hz
127 V AC, 50/60 Hz
120 V AC, 50/60 Hz
110 V AC, 50/60 Hz
100 V AC, 50/60 Hz
110 - 127 V AC, 60 Hz / 220 V AC, 50/60 Hz
Maximum Power Input 450 VAC
Ingress Protection (EN 60529) IPX4
Protection Class Class I
Electrical Motor Duty Cycle 2 minutes ON /18 minutes OFF
Accumulator Pb AKU 2 x 12 V / 1,2 Ah / Fuse 15A
Control Unit Fuse
Version 230 V
Version 127 V
Version 120 V
Version 110 V
Version 100 V
2 x T2.0A L 250 V
2 x T4.0A L 250 V
2 x T4.0A L 250 V
2 x T4.0A L 250 V
2 x T4.0A L 250 V
NOTE Upon request, LINET ® can deliver hospital beds with electrical specications that comply with regional standards (custom
voltage, dierent mains plugs).
5.4 Environment conditions (Multicare X)
Use Conditions
Ambient Temperature 10°C — 40°C
Relative Humidity 30% — 75 %
Atmospheric Pressure 795 hPa — 1060 hPa
Storage and Transport Conditions
Ambient Temperature -20°C — 50°C
Relative Humidity 20% — 90 %
Atmospheric Pressure 795 hPa — 1060 hPa
5.5 Electrical Specications (Multicare X)
Parameter Value
Lifting Pole Safe Working Load 75 kg
Maximum Mass of Mobile Hospital Bed (Empty Bed Maximum
Mass + Safe Working Load)
525 kg
Maximum Patient Weight for Application Environment 1, 2 in
accordance with IEC 60601-2-52
185 kg
Sound Pressure Level 50 dBA
Application Environment in accordance with IEC 60601-2-52 1, 2