1.6.3 Wash Label (mattress)
1.7 Acoustic signalisation (OptiCare)
SOUND MEANING
REPEATED BEEP: 0,2s sound / 1s silence
(high priority)
mattress ination fault (fault number 2)
mattress overpressure (fault number 5)
air valves fault (fault number 14)
CPR mattress deation fault (fault number 13)
BOD valves fault (fault number 6)
clogged air lter (fault number 8)
automatic deation is not available (Use manual CPR!)
REPEATED BEEP: 0,2s sound / 5s silence
(medium priority)
pressure sensor fault (fault number 4)
mattress disconnected
air leak
REPEATED BEEP: 0,2s sound / 10s silence
(low priority)
mattress deation fault (fault number 3)
MCM ventilator fault (fault number 7)