7 Use and Storage Conditions (OptiCare)
DANGER!
Danger to life due to electric shock!
To ensure the bed’s class I protection against electric shocks:
► Ground the mains.
► Use exclusively Hospital Grade or Hospital Only receptacles for grounding.
Eleganza 5 with OptiCare integrated mattress replacement system and Multicare with OptiCare integrated mattress replacement
system are designed for use in rooms for medical purposes. Electrical installations must therefore meet local norms laying down
the necessary conditions for electrical installations.
► Disconnect the bed from the mains in exceptional cases (i.e. lightnings, earthquake).
OptiCare integrated mattress replacement system is not suitable for indoor environments containing ammable gases
(except oxygen cylinders).
7.1 Storage
When SCU is not in use:
► Logg o the mattress.
► Unplug mains cable.
When mattress is not in use:
► Unclip all 5 air pipes.
► Undo webbing strap next to air pipes.
► Make sure mattress has been cleaned and is completely dry inside and out (see Cleaning/Disinfection).
► Deate mattress and leave air connector open (CPR position).
► Roll mattress up carefully to get air out completely.
► Place mattress in storage bag.
Store in a dry and safe place and keep away from sharp objects.
8 Scope of Delivery and Product Variants (OptiCare)
8.1 Delivery
► Upon receipt, check that the shipment is complete as specied on the delivery note.
► Notify the carrier and supplier of any deciencies or damages immediately as well as in writing or make a note on the
delivery note.
8.2 Scope of Delivery
■ OptiCare integrated mattress
■ SCU (System Control Unit)
■ Instructions for use