D9U003VS0-0101_01
33
OptiCare for Eleganza 5
10.3.4 MATTRESS PREPARED FOR PATIENT
When „MATTRESS INFLATION“ text disappears
and Mattress icon is fully green the mattress is ready for a patient.
NOTE Flashing Mattress Icon with selected OPT Mode
indicates continuing ination.
To use the mattress:
position patient on the mattress.
10.3.5 PATIENT ON THE MATTRESS (OPT MODE)
Integrated Micro-Climate Management system will start working
automatically with intensity set by MCM button when the patient
gets into bed and stop if the patient gets out. As long as
the patient remains on the mattress automatic optimization will
continue. Optimization will occur if the patient´s position changes
su󰀩ciently to trigger Optimization Detection or if initiated by
the Optimization automatic timer.
NOTE During pressure optimization the Mattress icon is ashing.
Fully green mattress icon indicates the mattress has achieved
optimum pressure.
Optimization will stop working and the mattress air pressure will be set at a xed level if
the mattress support platform is tilted by 10 and more degrees (Anti-Trendelenburg Tilt)
the mattress support platform is tilted by 10 and more degrees (Lateral Tilt)
the mattress support platform is tilted by 5 and more degrees (Trendelenburg Tilt)
tilt reduction of the mattress support platform has value of 7 or less degrees (Anti-Trendelenburg Tilt)
tilt reduction of the mattress support platform has value of 7 or less degrees (Lateral Tilt)
tilt reduction of the mattress support platform has value of 3 or less degrees (Trendelenburg Tilt)
In this case text „TILT NO OPT“ is displayed.
NOTE If at any time nursing sta󰀨 feel it necessary to re-optimize the patient then this can be initiated manually by touching the
button
. This does not over-ride Optimization settings and this process will continue as before.
Manual pressure optimization:
press button .
To set intensity of Micro-Climate Management:
press button
.
10.3.6 MAXIMUM MATTRESS INTERNAL PRESSURE
To set Maximum Mattress Internal Pressure:
press button until „MAX“ appears on the place
of the Mode Name.
NOTE During ination Mattress icon is ashing until it turns green.
NOTE During MAX mode text „MAX“ alternates with text indicating
remaining minutes of the MAX mode (e.g. 30´).
NOTE After 30 minutes the pressure optimization starts again.
Countdown is displayed on the place of Mode Name (9).
NOTE Maximum Mattress Internal Pressure can be operated with or without a patient on the mattress.
NOTE To extend Maximum Mattress Internal Pressure you can press Mode Button again during last 5 minites of the MAX Mode.
Fig. MATTRESS PREPARED FOR PATIENT
Fig. Patient On The Mattress
Fig. Maximum Mattress Internal Pressure