6 Technical Specication (OptiCare)
All technical data are rated data and are subject to construction and manufacturing tolerances.
WARNING!
If Eleganza 5 bed or Multicare bed are used with OptiCare integrated mattress replacement system, respect 
values of mechanical and electrical specications which can harm none of them!
6.1 Mechanical Specications
Parameter Value
External dimensions of Inated Mattress 214 cm x 86 cm x 22 cm
External dimensions of SCU 36 cm x 22 cm x 10 cm
Weight of Inated Mattress 10 kg (22 lbs)
Weight of SCU 6 kg (13.1 lbs)
Ination time after storage max. 15 min (typical < 10 min)
CPR deation time (depending on patient weight, chosen mode - optimization 
or maximum internal pressure - and on type of CPR - electric or manual)
max. 30 s
Maximum Mattress Load (SWL) 250 kg / 550 lbs / 39 stones
Minimum Mattress Load 40 kg
6.2 Environment conditions
Use Conditions
Ambient Temperature 10°C - 40°C
Relative Humidity 30% - 75 %
Atmospheric Pressure 795 hPa - 1060 hPa
Remain Inated in Transport Mode 12 hours (whem starting from Maximum Inate Mode)
Sound Pressure Level max. 45 dBa (normal operation without alarm)
Storage and Transport Conditions
Ambient Temperature -40°C - 70°C
Relative Humidity 10% - 100 % (non-condensing)
Atmospheric Pressure 795 hPa - 1060 hPa
6.3 Electrical Specications
Parameter Value
Input Voltage, Frequency:
EU version
USA version
220-240 V AC, 50/60 Hz
110-127 V AC, 50/60 Hz
Maximum Power Input max. 40 VA (when operating from mains supply)
Ingress Protection (EN 60529) IP44
Protection Class Class I with Applied Parts Type B
Electrical Safety in conformity with EN 60601-1
Fuse 2 x T1AH Anti-surge Fuse