4 Intended use (OptiCare)
The intended use is to prevent and support treatment of pressure ulcers.
OptiCare integrated mattress replacement system is compatible only with Eleganza 5 bed and Multicare bed.
4.1 User population
► Caregivers (nurses, doctors, technical personnel, transport personnel, cleaning personnel)
► Adult patients (weight >= 40 kg, height >= 146 cm, BMI >= 17) at moderate risk levels in the intensive care (Application
Environment 1 and 2, as in IEC 60601-2-52); Patients with any stage/category of pressure ulcers
4.2 Contraindications
OptiCare integrated mattress replacement system is contraindicated for patients:
► with cervical traction
► or unstable spinal fractures, spinal cord injury, fractures at risk of complication by a moving surface,
► or trauma patients where spinal injuries have not been excluded or cleared
► or prone position
4.3 Operator
► Caregiver