Mattress Detection System (MDS)
Sensors in the System Control Unit (SCU) Area A, B and ODV mattress air connectors detect that a valid air connector has been
connected. When all three correct air connectors are detected, the SCU will enter Standby mode. In Standby mode mattress areas
A and B are inflated to a static pressure ready for a patient to be placed onto the mattress and Optimization to start.
9.1 SCU (System Control Unit)
WARNING!
OptiCare mattress is compatible with System Control Unit delivered by manufacturer only!
► Do not use any other System Control Unit with OptiCare mattress!
CAUTION!
Material damage due to incorrect installation of SCU!
► If the SCU does not come factory-tted, have it installed by a service engineer authorised by LINET ®.
9.2 Replacing the mattress
When replacing the OptiCare mattress with an alternative active mattress from the LINET integrated systems, the system will auto-
matically detect that type of mattress that has been connected and switch to the correct iBoard Standard control screen. If replacing
the OptiCare mattress with one that is not from the LINET integrated systems range then you will need to cancel the Mattress Not
Connected alarm.
Logging out the OptiCare:
When replacing the OptiCare mattress by a mattress that is not part of the LINET OptiCare integrated mattress range, it is necessa-
ry to log out OptiCare.
► Press and hold
to log out mattress. Text „M OFF“ is displayed.
10 Manipulation (OptiCare for Eleganza 5)
Preparing OptiCare for patient
DANGER!
Risk of suocation due to air-impermeable mattress cover!
► Use mattress cover correctly.
► The nursing sta are responsible for the safety of the patient on the mattress cover.
WARNING!
Risk of injury when positioning patient on the bed!
Before positioning the patient on the bed:
► Ensure that mattress is completely and correctly inated.
► Ensure that mattress is correctly secured with safety straps.
CAUTION!
Material damage due to dampness or contamination!
► Make sure mattress cover has been cleaned and is completely dry (see Cleaning/ Disinfection).
Preparation
► Inate mattress.
► Put a sheet loosely on the mattress if not prescribed otherwise by qualied personnel.
Positioning the Patient on the Bed
► Put the patient on the mattress.
Create the ideal patient position:
► If additional blankets or sheets are used, make sure that ease of movement is sucient.
► Ensure that blankets, sheets, clothing etc. do not cause pressure injuries (e.g. due to creases, seams etc.).
► Do not place any additional sheets, blankets etc. between mattress and patient.