6.4 Electromagnetic Compatibility
OptiCare integrated mattress replacement system is intended for hospitals except for near active HF surgical equipment
and the RF shielded room of a medical system for magnetic resonance imaging, where the intensity of EM disturbances is
high.
OptiCare integrated mattress replacement system has dened no essential performance.
WARNING!
It is recommended to avoid the use of this device next to or in block with other device, because it could lead
to improper operation. If such use is needed, this device and the other equipment should be under surveillan-
ce to verify proper operation. (Does not apply for compatible medical bed from LINET)
List of used cables:
► Mains cable, maximum length 6 m
WARNING!
Use of the accessories, converters and other cables, than specied and provided by manufacturer of this
OptiCare integrated mattress replacement system could lead to increase of electromagnetic emission or lower
the electromagnetic immunity of this OptiCare integrated mattress replacement system and lead to improper
operation.
WARNING!
Mobile RF communication device (including end use devices like antenna cables and external antenna) should
not be used closer than 30 cm (12 inches) from any part of this OptiCare integrated mattress replacement
system, including cables specied by manufacturer. Otherwise this could lead to deterioration of functionality
of this OptiCare integrated mattress replacement system.
WARNING!
Do not overload the OptiCare integrated mattress replacement system (SWL), respect the duty cycle (INT.) and
consider chapter 12 Maintenance in order to maintain the basic safety with regard to electromagnetic distur-
bances for the expected service life.
6.4.1 Manufacturer instructions - electromagnetic emissions
Emission Test Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations / icker emissions
IEC 61000-3-3
Complies