OptiCare for Multicare
12.2 Automatic CPR (Multicare with OptiCare)
Cardiopulmonary Resuscitation Mode causes the mattres to de ate completely to facilitate resuscitation of patient. Typical de ation
time for OptiCare is 15 seconds (max. 30 s). After 55 minutes of CPR Mode there will be an audio alarm every 30 seconds. After 60
minutes system will return to OPT Mode or MAX In ate Mode.
Fig. CPR button
To activate the CPR Mode:
Press and hold CPR button on Multiboard for at least 3 s.
The mattress platform will straighten up.
The mattress de ates completely and mattress support platform will enter the CPR position.
NOTE: It is possible to use ACP also to activate CPR Mode.