Additional Activities
RET
eval
Device User Manual 24
Data folder to archive the results in machine-readable formats for programmatic
analysis.
Step 6. Delete results from the Reports directory to remove them from the device. If you are
saving results in multiple formats (e.g., PDF and JPEG), all formats have to be deleted in
order to remove the result from the device and make space for future tests. The raw
data files (.rff) and XML files (.rffx) do not need to be deleted. The device will
automatically remove those files as appropriate.
Updating firmware
Periodically LKC publishes an update to the device firmware. Firmware updates are provided at
no charge for the first year, afterward an annual support fee may be required for continued
support and firmware updates.
Follow these steps to update the device firmware:
Step 7. Download the firmware update file to the PC. (Follow instructions in the firmware
update notice to find and download the update.)
Step 1. Connect the USB cable to the PC.
Step 2. Place the device into the docking station.
Step 3. Wait for the device to appear as an external drive on the PC.
Step 4. Copy the firmware update file from the directory on the PC to the Firmware directory
on the device.
Step 5. Eject the external drive that represents the device from the PC.
Step 6. Remove the device from the docking station.
Step 7. Select Settings, then System, then Change Settings, then Update Firmware.
Step 8. Select the firmware update you want.
Step 9. Select Next.
Step 10. Wait while the firmware is updated.
Step 11. After the firmware update completes, the device will restart automatically.
If the RET
eval
fails during the firmware update, verify that the firmware update file was
downloaded and copied to the device correctly by repeating steps 5 through 12.
Electronic medical record (EMR) support
The RET
eval
device supports EMR integration through the use of passing files between a host PC
and the EMR folder on the RET
eval
device. The patient ID and birthdate can be electronically
transferred to the device, and only need to be confirmed on the device before starting a test. On
completion of a test, docking the RET
eval
device back with the PC enables results to be
electronically pulled off the device and into the EMR. Contact LKC for more details about
currently supported EMR systems and integration options with your EMR.