Regulatory and Safety Information
RET
eval
Device User Manual 92
0 % UT; 250/300
cycle for 50 Hz and
60 Hz, respectively
Single phase: at 0°
0 % UT; 250/300
cycle for 50 Hz and
60 Hz, respectively
Single phase: at 0°
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
Power frequency magnetic fields
should be that of a typical commercial,
hospital, or home environment.
Guidance and Manufacturer’s Declaration – Immunity
The RET
eval
device is intended for use in the electromagnetic environment specified below.
The customer or user of the RET
eval
device should ensure that it is used in such an environment.
Electromagnetic Environment –
Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V, 0.15 MHz – 80
MHz
6 V in ISM radio
bands between 0.15
MHz and 80 MHz
80 % AM at 1 kHz
3 V/m Professional
80 MHz – 2,7 GHz
80 % AM at 1 kHz
Table 9 of IEC
60601-1-2:2014
Portable and mobile
communications equipment should
be separated from the RET
eval
device by no less than the distances
calculated/listed below:
D=
3.5
V1
√
P, 150kHz to 80MHz
D=
3.5
E1
√
P, 80 to 800 MHz
D=
7
E1
√
P, 800 MHz to 2.5 GHz
where P is the max power in watts
and D is the recommended
separation distance in meters. Field
strengths from fixed transmitters, as
determined by an electromagnetic
site survey, should be less than the
compliance levels (V1 and E1).
Interference may occur in the
vicinity of equipment containing a
transmitter.
To assure continued effectiveness,
only use cables and accessories
supplied by LKC which are
specifically designed for use with
the RET
eval
device.