4. Close the Lid, check if the Cal tool is placed
properly, no bending.
5. Move cursor to “2.Ver2-[a] Dark Paper” and
double press “ ”. After the screen shows a black
arrow beside the dimension value, wait for about 5
seconds.
6. Once a green mark appears on the screen, first
calibration is completed.
Step B: Transmission Calibration
1. Open lid and remove calibration tool A (1.0mm For
IR).
2. Insert the Calibration tool B (1.2mm for IR), side
with foam downwards.
3. Adjust the position of the Calibration tool to the
center part of the bill path.
Cal tool is not bowed
Make sure the tool is placed
at center.