4. Close the Lid, check if the Cal tool is placed
properly, no bending.
5. Move cursor to “3. Ver2-[b] Bright Paper” and
double press “ ”. After the screen shows a black
arrow beside the dimension value, wait for about 5
seconds.
6. Once a green mark appears, second calibration is
completed.
7. Open lid and remove calibration tool B (1.2mm for
IR).
8. Move to “Exit” and press “ ” to go back to upper
layer.
4.4.2 Adjust MG
To adjust the retrieved value of magnetic sensor.
1. Highlight the option and press “ ”.
2. Adjust the value of magnetic sensor by pressing
the keys “
F1” and “F2”. (from 000 to 128)
Note: To adjust the MG value, please contact MA.
4.4.3 Initialization
Initializing the sensor.
Cal tool is not bowed