Service Manual NC-6000 Date: OCT 1
st
, 2012 (Version 1.0)
Masterwork Automodules Tech. Corp., Ltd
Confidential
- 22 -
4. Close the Lid, check if the Cal tool is placed
properly, no bending.
5. Move cursor to “3. Ver2-[b] Bright Paperand
double press “ . After the screen shows a black
arrow beside the dimension value, wait for about 5
seconds.
6. Once a green mark appears, second calibration is
completed.
7. Open lid and remove calibration tool B (1.2mm for
IR).
8. Move to Exitand press to go back to upper
layer.
4.4.2 Adjust MG
To adjust the retrieved value of magnetic sensor.
1. Highlight the option and press “ ”.
2. Adjust the value of magnetic sensor by pressing
the keys “
F1” and “F2”. (from 000 to 128)
Note: To adjust the MG value, please contact MA.
4.4.3 Initialization
Initializing the sensor.
Cal tool is not bowed