Service Manual – NC-6000 Date: OCT 1
st
, 2012 (Version 1.0)
Masterwork Automodules Tech. Corp., Ltd
Confidential
- 20 -
4.4 Calibration mode
Before calibrating sensors.
1. Highlight the option and press “ ”
2. Use “F1” and “F2” key to adjust the dimension
value of calibration paper.
Note: Dimension value is stated on each calibraiton
paper.
4.4.1 Calibrate sensors
To calibrate the IR image.
This function does not exist in NC6000.
Step A: Transmission Calibration.
1. Open the lid and insert the Calibration tool A
(1.0mm for IR), side with foam downwards.