SedLine Sedation Monitor Chapter 5: Menu Options
www.masimo.com 38 Masimo
PSI
About
An informational read-only screen with definition of PSI:
The Patient State Index (PSI) is related to the effect of certain anesthetic agents on a patient.
Alarms
Options Description
Factory
Default
Selectable
Settings
High Limit Upper limit that triggers an alarm 50 Off, 5 to 99
Low Limit Lower limit that triggers an alarm 25 Off, 1 to 95
High Caution
Range
Value below High Limit to trigger
caution light
Off 0 to 10
Low Caution
Range
Value above Low Limit to trigger caution
light
Off 0 to 10
Silence
Duration
Length of time that the audible alarm
remains silenced
2 min
30 sec, 1, 2, or 5
min
Audible Alarms Disables audible alarm On On or Off