User Manual MAS-100 Regulus en.docx, Version: 4.0
12) After successful login, the PC Software starts in the Adjustment/Calibration menu.
13) Select the perforated lid of the sampling head:
300×0.6 or 400×0.7
14) Select Automatic adjustment.
15) Set the adjustment parameters:
Adjustment flow
Adjustment span
Start the process with the button.
Pressing afterwards aborts the process.
Note
The standard factory setting for the Adjustment flow is 100SLPM.
Note
The standard factory setting for the Adjustment span is 10SLPM.
16) Press Yes to proceed.
Do not add or remove instruments.
Pressing No aborts the process.