Page 106 of 188
User Manual MAS-100 Regulus en.docx, Version: 4.0
17) The air sampler will now adjust through all three adjustment points. No action from
the user is required, the process will continue automatically.
18) Press the Yes button to accept the adjustment and
store the values in the air sampler.
Pressing No aborts the process.
19) If used, remove the MAS-100 Family Calibration Disk or the Petri dish.
20) End of process.