12) After successful login, the PC Software starts in the Adjustment/Calibration menu.
13) Select the head and the perforated lid on it:
300 × 0.6 or 400 × 0.7
14) Choose the calibration validity:
– Values from 0 to 12 months may be entered.
– A value of 0 disables the calibration validity
check upon starting up the air sampler.
15) Activate the optional function Measurement block-
ing upon expired calibration, if desired.
16) Set the absolute Maximum allowable deviation:
2.5 % to 25.0 %
17) Start the process with the button.
Pressing afterwards aborts the process.
Note
The standard factory setting for the Calibration validity is 12 months.
Note
Since PC software V3.1.14, the Disable calibration validity function is only ac-
tive for Factory user level.